Figure 3-3: Basic Pneumatic Connections Flow Diagram - Teledyne 300E Instruction Manual

Carbon monoxide analyzer
Table of Contents

Advertisement

Model 300E Carbon Monoxide Analyzer
Instruction Manual
4.
Once the appropriate pneumatic connections have been made, check all
pneumatic fittings for leaks using a procedure similar to that defined in
Section 9.3.3.
Figure 3–3: Basic Pneumatic Connections Flow Diagram
The Inlet/Outlet labels appearing in the various pneumatic diagrams found in this
manual (such as Figure 3–3).
The order in which they appear in the illustrations IS NOT necessarily the same order
in which they are arranged on the rear panel of the instrument.
P/N 04288 REV: A
Calibrated
CO Gas
MODEL 701
VENT
Zero Air Generator
Calibrated
CO Gas
Valve
MODEL 701
Zero Air Generator
Sample Gas pressure must equal ambient atmospheric pressure.
In applications where the sample gas is received from a pressurized
manifold, a vent must be placed as shown to equalize the sample gas
with ambient atmospheric pressure before it enters the analyzer.
This vent line must be:
At least 0.2m long
q
No more than 2m long
q
Vented outside the shelter or immediate area surrounding the
q
instrument
No Valve Options Installed
MODEL 700
Gas Dilution
Calibrator
Sample
Pressure Span
Vent Span
IZS
Exhaust
OR
Needle Valve to
control Flow
Sample
Pressure Span
Vent Span
VENT
IZS
Exhaust
NOTE
CAUTION
Getting Started
Source of
SAMPLE Gas
Removed
during
Calibration
MODEL
300E
Source of
SAMPLE Gas
Removed
during
Calibration
MODEL
300E
29