Figure 2-7. Gas Line Connections from Calibrator – Basic Configuration
For the analyzer's basic configuration, in addition to tubing, pressure, venting, and
exhaust requirements set out in Section 2.3.2.1, attach the following pneumatic lines:
S
AMPLE
Connect a sample gas line to the SAMPLE inlet.
C
ALIBRATION
NO
CAL GAS & ZERO AIR SOURCES: The source of calibration gas is attached to
2
the SAMPLE inlet, but only when a calibration operation is actually being performed.
V
ENTING
Vent the output of the calibrator if calibrator not already vented.
E
XHAUST
Attach an exhaust line to the EXHAUST outlet fitting and vent outside the shelter.
24
Teledyne API N500 CAPS NO