Figure 3-3: Basic Pneumatic Connections Flow Diagram - Teledyne 400E Instruction Manual

Ozone analyzer
Table of Contents

Advertisement

Model 400E Ozone Analyzer
Instruction Manual
MODEL 701
Zero Air Generator

Figure 3-3: Basic Pneumatic Connections Flow Diagram

3. The exhaust from the pump and vent lines should be vented to atmospheric
pressure using maximum of 10 meters of ¼" PTEF tubing. Venting should be
outside the shelter or immediate area surrounding the instrument.
The Inlet/Outlet labels appearing in the various pneumatic
diagrams found in this manual (such as Figure 3-3) are
The order in which they appear in the illustrations IS NOT
necessarily the same order in which they are arranged on the
P/N 04316 Rev: B
Gas Dilution Calibrator
(w/ Photometer Option)
Venting should be outside the shelter or immediate area
surrounding the instrument and conform to all safety
requirements regarding exposure to O
arranged to make the illustration as clear as possible.
rear panel of the instrument.
No Valve Option Installed
MODEL 700
or
MODEL 401
Ozone Generator
VENT
Sample
Exhaust
Span
Zero Air
Dry Air
CAUTION
NOTE
Getting Started
Source of
SAMPLE GAS
Removed during
calibration
MODEL
400E
.
3
31