Making Measurements
Using Markers

To Set Measurement Parameters Using Markers

The analyzer allows you to set measurement parameters with the markers, without going through the usual
key sequence. You can change certain stimulus and response parameters to make them equal to the current
active marker value.
Setting the Start Frequency
1. Press
Marker Fctn
position the marker at the value that you want for the start frequency.
2. Press
MARKERSTART
Figure 1-22 Example of Setting the Start Frequency Using a Marker
Setting the Stop Frequency
1. Press
Marker Fctn
position the marker at the value that you want for the stop frequency.
2. Press
MARKERSTOP
1-34
and turn the front panel knob, or enter a value from the front panel keypad to
to change the start frequency value to the value of the active marker.
and turn the front panel knob, or enter a value from the front panel keypad to
to change the stop frequency value to the value of the active marker.