Operating Concepts

Knowing the Instrument Modes

Knowing the Instrument Modes

There are five major instrument modes of the analyzer:

• network analyzer mode

• external source mode

• tuned receiver mode
• frequency offset operation
• harmonic mode operation (Option 002)
Network Analyzer Mode
This is the standard mode of operation for the analyzer, and is active after you press
or switch on
Preset
the AC power. This mode uses the analyzer's internal source.
Pressing
returns the analyzer to the
System
INSTRUMENT MODE
NETWORK ANALYZER
"normal" network analyzer operating mode.
External Source Mode
This mode allows the analyzer to phase lock to an external CW signal. External source mode is best used for
unknown signals, or for signals that drift. If a synthesized external source is used, the tuned receiver mode is
recommended because it is faster.
Primary Applications
External source mode is useful in several applications:
• when your test device is a mixer or other frequency translation device
• in automated test applications where a source is already connected to the system, and you do not want
to switch between the system source and the analyzer's internal source.
Typical Test Setup
A typical test setup using the external source mode is shown in
Figure
7-46. The same test setup is
applicable for either manual or automatic external source mode operation.