Figure 2-16
Connections for Power Meter Calibration
3. Select the analyzer as the system controller:
Local
SYSTEM CONTROLLER
4. Set the power meter's address:
SET ADDRESSES
ADDRESS: P MTR/GPIB
5. Select the appropriate power meter by pressing
displayed (436A or 438A/437).
NOTE
The E4418B and E4419B power meters have a "437 emulation" mode. This allows these
power meters, with an 848X-series power sensor, to be used with the network analyzer. In
this step, when selecting a power meter, choose the 438A/437 selection.
6. Press
Cal
PWRMTR CAL
the correction factors as listed on the power sensor. Press
the frequency in MHz)
for each correction factor. When finished, press
DONE
x1
7. Perform a one sweep power meter calibration over the IF frequency range at 5 dBm:
ONE SWEEP
5
x1
TAKE CAL SWEEP
High Dynamic Range Swept RF/IF Conversion Loss
(where aa is the power meter GPIB address)
aa
LOSS/SENSR LISTS
M/
CAL FACTOR
until the correct model number is
POWER MTR [ ]
CAL FACTOR
ADD
FREQUENCY
(where nnn is the calibration factor number)
nnn
DONE
Making Mixer Measurements
x1
and enter
SENSOR A
(where fff is
fff
.
RETURN
2- 21