Making Mixer Measurements
Conversion Compression Using the Frequency Offset Mode
CAUTION
To prevent connector damage, use an adapter (part number 1250-1462) as a connector saver
for R CHANNEL IN.
Figure 2-28
Connections for the First Portion of Conversion Compression Measurement
6. To view the absolute input power to the analyzer's R channel, press:
Meas
INPUT PORTS
7. To store a trace of the receiver power versus the source power into memory and view data/memory,
press:
Display
DATAMEM
DATA/MEM
This removes the loss between the output of the mixer and the input to the receiver, and provides a linear
power sweep for use in subsequent measurements.
8. Make the connections as shown in
CAUTION
To prevent connector damage, use an adapter (part number 1250-1462) as a connector saver
for R CHANNEL IN.
2-40
R
Figure
2-29.