displayed (436A or 438A/437).
NOTE
The E4418B and E4419B power meters have a "437 emulation" mode. This allows these
power meters, with an HP/Agilent 848X-series power sensor, to be used with the network
analyzer. In this step, when selecting a power meter, choose the 438A/437 selection.
6. Set test port power to the approximate desired corrected power.
7. Press
Cal
PWRMTR CAL
test device. For example, if you enter
8. If you want the analyzer to make more than one power measurement at each frequency data point,
press:
NUMBER OF READINGS
If you increase the number of readings, the power meter correction time will substantially increase.
9. Press
PWRMTR CAL
Cal
NOTE
Because power meter calibration requires a longer sweep time, you may want to reduce the
number of points before pressing
calibration is finished, return the number of points to its original value and the analyzer will
automatically interpolate this calibration. Some accuracy will be lost for the interpolated
points.
The analyzer will use the data table for subsequent sweeps to correct the output power level at each
measurement point. Also, the status annunciator PC will appear on the analyzer display.
NOTE
You can abort the calibration sweep by pressing
10. Remove the power sensor from the analyzer test port and connect your test device.
Calibrating for Increased Measurement Accuracy
and enter the test port power level that you want at the input to your
, the display will read CAL POWER 10.
10
x1
, (where n = the number of desired iterations).
n
x1
ONE SWEEP
TAKE CAL SWEEP
TAKE CAL SWEEP
Power Meter Measurement Calibration
.
. After the power meter
PWRMTR CAL OFF
.
6- 37