the left of the screen, indicating that error correction is on.

Enhanced Reflection Calibration

19. If you are measuring a bilateral device and want to remove the load match error, activate the enhanced
reflection calibration by pressing:
ENHANCED RESPONSE
ENH. REFL. on OFF
NOTE
You can save or store the measurement correction to use for later measurements. Refer to
Chapter 4 , "Printing, Plotting, and Saving Measurement Results"
20. This completes the enhanced response correction procedure. You can connect and measure your device
under test.
Enhanced Frequency Response Error Correction
until ON is selected
Calibrating for Increased Measurement Accuracy
for procedures.