Calibrating for Increased Measurement Accuracy
Full Two-Port Error Correction (ES Analyzers Only)
17. To compute the error coefficients, press:
DONE 2-PORT CAL
The analyzer displays the corrected measurement trace. The analyzer also shows the notation Cor at
the left of the screen, indicating that error correction is on.
NOTE
You can save or store the measurement correction to use for later measurements. Refer to
Chapter 4 , "Printing, Plotting, and Saving Measurement Results"
18. This completes the full two-port correction procedure. You can connect and measure your device under
test.
6-32
for procedures.