Polar Format - Agilent Technologies 8753ET User Manual

Network
Hide thumbs Also See for 8753ET:
Table of Contents

Advertisement

Operating Concepts
Analyzer Display Formats

Polar Format

The
softkey displays a polar format as shown in
Figure
7-10. Each point on the polar format
POLAR
corresponds to a particular value of both magnitude and phase. Quantities are read vectorally: the
magnitude at any point is determined by its displacement from the center (which has zero value), and the
phase by the angle counterclockwise from the positive x-axis. Magnitude is scaled in a linear fashion, with
the value of the outer circle usually set to a ratio value of 1. Since there is no frequency axis, frequency
information is read from the markers.
The default marker readout for the polar format is in linear magnitude and phase. A log magnitude marker
and a real/imaginary marker are available in the polar marker menu.
Figure 7-10
Polar Format
7- 29