Agilent Technologies 8753ET User Manual page 272

Network
Hide thumbs Also See for 8753ET:
Table of Contents

Advertisement

Printing, Plotting, and Saving Measurement Results
Saving Measurement Results
Data Arrays
Press
Save/Recall
DEFINE DISK-SAVE
Data created the first time in this manner will be saved as filename "DATA00.d1." The file extension .d1
indicates that the data from the analyzer's channel 1 is error corrected data only if the analyzer's error
correction feature is enabled (in other words, you have performed a calibration). Otherwise, the data is the
same as data stored in the analyzer's raw data arrays. Data stored in the data arrays does not have any
formatting applied to it.
Format Arrays
Press
DEFINE DISK-SAVE
Save/Recall
Data created the first time in this manner will be saved as filename "FILE00.f1". The file extension .f1
indicates the data is formatted per
you've selected, data in the format arrays includes data in the data arrays plus one or more of the following
features:
• Trace math (i.e data-memory)
• Gating (Option 010)
• Electrical delay
• Conversion (for complex impedance (Z), admittance (Y), etc.)
• Transform (Option 010)
• Format (log, lin, phase, delay, SWR, excluding Smith and Polar)
• Smoothing
In each of these examples, most users will select
DEFINE DISK-SAVE
created with file extension .g0. This is a Hewlett-Packard Graphics Language (HPGL) file.
4-48
DATA ARRAY ON
FORMAT ARY ON
Figure 4-13
using the analyzer's channel 1. Depending on what features
SAVE USING ASCII
softkey menu. If
GRAPHICS on OFF
.
, under the
is turned ON, an additional file will be