Figure 6-28 Two-Port Cal Set 2
7. Connect the ECal module between adapter A3 and adapter A2.
8. Press
Cal
ECal MENU
9. Press
FULL 2-PORT
10. Save the results to disk. Name the file "PORT2."
11. Determine the electrical delay of adapter A3. If you have adapter specifications that identify the electrical
delay, you may use that information and continue with
know the delay of adapter A3, perform the "Determine the Electrical Delay" procedure on
Adapter Removal Using ECal (ES Analyzers Only)
.
to perform the second 2-port error correction using the ECal module.
Calibrating for Increased Measurement Accuracy
"Remove the Adapter" on page
6-73. If you do not
page
6-72.
6- 71