Measuring Magnitude And Insertion Phase Response; Measuring The Magnitude Response - Agilent Technologies 8753ET User Manual

Network
Hide thumbs Also See for 8753ET:
Table of Contents

Advertisement

Measuring Magnitude and Insertion Phase Response

This measurement example shows you how to measure the maximum amplitude of a surface acoustic wave
(SAW) filter and then how to view the measurement data in the phase format, which provides information
about the phase response.

Measuring the Magnitude Response

1. Connect your test device as shown in
Figure 1-2 Device Connections for Measuring a Magnitude Response
2. Press
and choose the measurement settings. For this example the measurement parameters
Preset
are set as follows:
Meas
Trans:FWD S21 (B/R)
Center
134
Span
50
M/
3
Power
AUTO SCALE
Scale Ref
Chan 2
Meas
Trans:FWD S21 (B/R)
Scale Ref
AUTO SCALE
You may also want to select settings for the number of data points, averaging, and IF bandwidth.
3. Remove the device and connect the power cables together (thru) and perform a response calibration
using the following key presses.
Press
Chan 1
Cal
If the channels are coupled (the default condition), this calibration is valid for both channels.

Measuring Magnitude and Insertion Phase Response

Figure
1-2.
or on ET models:
M/
x1
or on ET models:
CALIBRATE MENU
TRANSMISSN
TRANSMISSN
RESPONSE
THRU
Making Measurements
.