Setting the Frequency Range
To set the center frequency to 134 MHz, press:
Center
134
M/
To set the span to 30 MHz, press:
Span
30
M/
NOTE
You could also press the
start frequency and stop frequency values.
Setting the Source Power
To change the power level to 5 dBm, press:
5
Power
x1
NOTE
You could also press
the power ranges to keep the power setting within the defined range.
Setting the Measurement
To change the number of measurement data points to 101, press:
Sweep Setup
NUMBER OF POINTS
To select the transmission measurement, press:
Meas
Trans: FWD S21 (B/R)
To view the data trace, press:
AUTOSCALE
Scale Ref

Step 3. Perform and apply the appropriate error-correction.

Refer to the
Chapter 5 , "Optimizing Measurement Results,"
errors.
To save the instrument state and error-correction in the analyzer internal memory, press:
SELECT DISK
Save/Recall

Step 4. Measure the device under test.

Replace any standard used for error-correction with the device under test.
To measure the insertion loss of the bandpass filter, press:
Marker Search
SEARCH: MAX
and
keys and enter the frequency range limits as
Start
Stop
POWER RANGE MAN
or on ET models:
TRANSMISSN
INTERNAL MEMORY
Making Measurements
Making a Basic Measurement
POWER RANGES
for procedures on correcting measurement
RETURN
SAVE STATE
and select one of