Setting The Cw Frequency - Agilent Technologies 8753ET User Manual

Network
Hide thumbs Also See for 8753ET:
Table of Contents

Advertisement

Making Measurements
Using Markers
Setting the Electrical Delay
This feature adds phase delay to a variation in phase versus frequency, therefore it is only applicable for
ratioed inputs.
1. Press
Format
PHASE
2. Press
Marker Fctn
position the marker at a point of interest.
3. Press
MARKERDELAY
to compensate for the phase slope at the active marker position. This effectively flattens the phase trace
around the active marker. You can use this to measure the electrical length or deviation from linear
phase.
Additional electrical delay adjustments are required on devices without constant group delay over the
measured frequency span.
Figure 1-27 Example of Setting the Electrical Delay Using a Marker

Setting the CW Frequency

1. To place a marker at the desired CW frequency, press:
and either turn the front panel knob or enter the value, followed by a unit terminator.
Marker
2. Press
Seq
SPECIAL FUNCTIONS
You can use this function to set the marker to a gain peak in an amplifier. After pressing
MKRCW FREQ
increasing input power.
1-38
.
and turn the front panel knob, or enter a value from the front panel keypad to
to automatically add or subtract enough line length to the receiver input
MKRCW
, activate a CW frequency power sweep to look at the gain compression with
.

Advertisement

Table of Contents
loading

This manual is also suitable for:

8753es

Table of Contents