Operating Concepts
Knowing the Instrument Modes
relative power of the measured harmonic in dBc. You must uncouple channels 1 and 2 for this measurement,
using the
COUPLED CHAN ON off
After uncoupling channels 1 and 2, you may want to change the fundamental power and see the resultant
change in relative harmonic power (in dBc).
both channels simultaneously, even though they are uncoupled in all other respects.
Frequency Range
The frequency range is determined by the upper frequency range of the instrument or system (3 or 6 GHz)
and by the harmonic being displayed. The 6 GHz operation requires an 8753ET/ES Option 006.
shows the highest fundamental frequency for maximum frequency and harmonic mode.
Table 7-4 Maximum Fundamental Frequency Using Harmonic Mode
Harmonic
Measured
8753ET/ES with Option 002
2nd Harmonic
3rd Harmonic
Accuracy and Input Power
Refer to the 8753ES and the 8753ET specifications and characteristics chapters of the reference guide for
recommendations on the maximum recommended input power and maximum source power.
Using power levels greater than the recommended values may cause undesired harmonics in the source and
receiver. The recommended power levels ensure that these harmonics are less than 45 dBc. Use test port
power to limit the input power to your test device.
7- 86
softkey set to OFF to allow alternating sweeps.
COUPLE PWR ON off
Maximum Fundamental Frequency
8753ET/ES with Option 002 and Option 006
1.5 GHz
1.0 GHz
allows you to change the power of
Table 7-4
3.0 GHz
2.0 GHz