Making Measurements
Using Markers
Figure 1-29
Example of Searching for the Minimum Amplitude Using a Marker
Searching for a Target Amplitude
1. Press
Marker Search
2. Press
SEARCH: TARGET
3. If you want to change the target amplitude value (default is 3 dB), press
value from the front panel keypad. You may also press
the new value.
4. If you want to search for multiple responses at the target amplitude value, press
and
SEARCH RIGHT
Figure 1-30 Example of Searching for a Target Amplitude Using a Marker
1-40
to access the marker search menu.
to move the active marker to the target point on the measurement trace.
.
TARGET
TARGET VALUE
Marker Search
and enter the new
to enter
SEARCH LEFT