Agilent Technologies 8753ET User Manual page 354

Network
Hide thumbs Also See for 8753ET:
Table of Contents

Advertisement

Calibrating for Increased Measurement Accuracy
Calibrating Using Electronic Calibration (ECal)
network analyzer display shows the non-random crosstalk data.
NOTE
If you are performing an ECal using two modules, selecting either option sets up the option
to be performed with both modules.
To Select the Manual Thru Calibration Option
1. To select the manual thru calibration option, press
2. Press
MAN'L THRU on OFF
3. Press
.
RETURN
4. Continue with step 2 of
To Select the Isolation Calibration Option
1. To select the isolation calibration option, press
2. Press
OMIT ISOLATION ON off
The isolation measurement is normally off. Therefore, the default setting for the
OMIT ISOLATION ON off
(
OMIT ISOLATION on OFF
3. Press
ISOLATION AVERAGES
keypad, and then press
NOTE
The isolation averaging default is set to take 10 sweeps. This isolation averaging value is
less than the network analyzer default sweep averaging value of 16.
4. Press
.
RETURN
5. Continue with step 2 of
6-60
until ON is selected.
"Perform the Calibration" on page
until OFF is selected.
softkey is ON. When isolation is not omitted
), isolation standards are measured.
, enter the isolation averages numeric value on the front panel
.
x1
"Perform the Calibration."
Cal
ECal MENU
6-61.
Cal
ECal MENU
CONFIGURE [ ]
.
CONFIGURE [ ]
.

Advertisement

Table of Contents
loading

This manual is also suitable for:

8753es

Table of Contents