Optimizing Measurement Results
Increasing Measurement Accuracy
Table 5-2 Differences between PORT EXTENSIONS and ELECTRICAL DELAY
Main Effect
Measurements Affected
Electrical Compensation
You can activate a port extension by pressing
. Then enter the delay to the reference plane.
EXTENSIONS ON
5-6
PORT EXTENSIONS
The end of a cable becomes the test
port plane for all S-parameter
measurements.
All S-parameters.
Intelligently compensates for 1
times or 2 times the cable's
electrical delay, depending on which
S-parameter is computed.
Cal
MORE
ELECTRICAL DELAY
Compensates for the electrical
length of a cable. Set the cable's
electrical length 1 for
transmission. Set the cable's
electrical length 2 for reflection.
Only the currently selected
S-parameter.
Only compensates for electrical
length.
PORT EXTENSIONS