Figure 6-8 Standard Connections for a One Port Reflection Error Correction
8. To measure the standard, when the displayed trace has settled, press:
NOTE
If the calibration kit that you selected has a choice between male or female calibration
standards, remember to select the sex that applies to the test port and not the standard.
The analyzer displays WAIT - MEASURING CAL STANDARD during the standard measurement. The
analyzer underlines the
9. Disconnect the open, and connect a short circuit to the test port.
10. To measure the standard when the displayed trace has settled, press:
SHORT
The analyzer measures the short circuit and underlines the
11. Disconnect the short, and connect an impedance-matched load to the test port.
12. When the displayed trace has settled, press:
, select the type of load you are using, and then press
LOADS
has finished measuring the load.
Notice that the
LOADS
softkey after it measures the calibration standard.
OPEN
softkey is now underlined.
Calibrating for Increased Measurement Accuracy
One-Port Reflection Error Correction
OPEN
softkey.
SHORT
DONE: LOADS
when the analyzer
6- 27