Table 6-2 Purpose and Use of Different Error Correction Procedures (Continued)
Correction Procedure
S
1-port
11
S
1-port
22
(ES Analyzers Only)
Full 2-port
(ES Analyzers Only)
*
*
TRL
/LRM
(ES Analyzers Only)
NOTE
Response calibration is not as accurate as other calibration methods.
Procedures for Error Correcting Your Measurements
Corresponding
Measurement
Reflection of any one-port
device or well terminated
two-port device.
Reflection of any one-port
device or well terminated
two-port device.
Transmission or reflection
of highest accuracy for
two-port devices.
Transmission or reflection
in noncoaxial environment
such as in a fixture or
on-wafer.
Calibrating for Increased Measurement Accuracy
Errors Corrected
Directivity, source match,
frequency response.
Directivity, source match,
frequency response.
Directivity, source match,
load match, isolation,
frequency response,
forward and reverse.
Directivity, isolation,
frequency response.
(forward and reverse)
Standard Devices
Short, open and load or
ECal module.
Short, open and load or
ECal module.
Short, open, load and thru
(2 loads for isolation) or
ECal module
Thru, reflect, line, or line,
reflect, match, or thru,
reflect, match.
6- 11