Making Measurements
Using Ripple Limits to Test a Device
frequency band may be changed to display the value of each band. To change the displayed frequency band
value, from the Ripple Test Menu press
and then use the
and
keys
RIPL VALUE BAND
(or the numerical keypad) to select the desired frequency band.
Viewing the Ripple Value in Absolute Format
When
is selected, the absolute ripple value of the selected frequency
RIPL VALUE [ABSOLUTE ]
band is displayed. The absolute ripple value is the measured maximum level minus the measured minimum
level within the frequency band. This value is displayed in dB.
Figure 1-71
shows the ripple test with absolute ripple value displayed for Frequency Band 1. The B1
indicates that the ripple value displayed is for Frequency Band 1. Notice that Frequency Band 1 passes the
ripple test. It has an absolute ripple value of 1.675 dB while the maximum ripple value entered for Frequency
Band 1 was 2.0 dB. Thus, even though the ripple test fails because of Frequency Band 3, the ripple passes in
Frequency Band 1.
Figure 1-71 Filter Pass Band with Absolute Ripple Value for Band 1 Activated
1-92