Figure 6-15 Two-Port Cal Set 1
4. Perform a full 2-port calibration between ports 1 and 2 using calibration standards appropriate for the
connector type at port 1 (the connector type for adapter A1). Save the calibration by selecting
Save/Recall
SAVE STATE
5. Connect adapter A3 to adapter A1 on port 1 as shown in
Figure 6-16 Two-Port Cal Set 2
6. Perform a full 2-port calibration between ports 1 and 2 using calibration standards appropriate for the
connector type at port 2 (the connector type for adapter A2). Save the calibration by selecting
Save/Recall SAVE STATE
NOTE
In the following steps, calibration data is recalled, not instrument states.
7. Press
Cal
MORE
. Name the file "PORT1."
. Name the file "PORT2."
ADAPTER REMOVAL
Calibrating for Increased Measurement Accuracy
Calibrating for Noninsertable Devices
Figure
6-16.
RECALL CAL SETS
.
6- 43