Figure 2-33
Connections for a Response Calibration
5. Perform a response calibration by pressing
NOTE
A full 2-port calibration will increase the accuracy of isolation measurements. Refer to
Chapter 5 , "Optimizing Measurement Results."
6. Make the connections as shown in
CAUTION
To get an accurate assessment of the LO-IF isolation, the proper LO power level must be
input to the LO port.
Figure 2-34
Connections for a Mixer Isolation Measurement
7. To adjust the display scale, press
The measurement results show the mixer's LO to RF isolation.
Cal
CALIBRATE MENU
Figure
2-34.
Scale Ref
AUTO SCALE
Making Mixer Measurements
Isolation Example Measurements
RESPONSE
.
.
THRU
2- 45