Agilent Technologies 8753ET User Manual page 427

Network
Hide thumbs Also See for 8753ET:
Table of Contents

Advertisement

leads to a menu for constructing a label for the user-modified cal kit. If a label is
LABEL KIT
supplied, it will appear as one of the five softkey choices in the select cal kit menu. The approach is
similar to defining a display title, except that the kit label is limited to ten characters.
TRL/LRM OPTION
KIT DONE (MODIFIED)
are defined and all classes are specified. Be sure to save the kit with the
it is to be used later.
Define Standard Menus
Standard definition is the process of mathematically modeling the electrical characteristics (delay,
attenuation, and impedance) of each calibration standard. These electrical characteristics (coefficients) can
be mathematically derived from the physical dimensions and material of each calibration standard, or from
its actual measured response. The parameters of the standards can be listed in
Table 7-1 Standard Definitions
a
System Z
= __________________________________________
0
__________________________
Disk File Name:
b
c
STANDARD
C0
C1
15
10
10
F
F/Hz
NO.
TYPE
1
2
3
4
5
6
7
8
a. Ensure system Z
of network analyzer is set to this value.
0
b. Open, short, load, delay/thru, or arbitrary impedance.
c. Open standard types only.
d. Load or arbitrary impedance only.
e. Arbitrary impedance only, device terminating impedance.
Each standard must be identified as one of five "types": open, short, load, delay/thru, or arbitrary
impedance.
brings up the TRL Option menu.
terminates the calibration kit modification process, after all standards
c
c
c
C2
C3
FIXED
27
36
45
10
10
SLIDING
or
2
3
F/Hz
F/Hz
OFFSET
Calibration Kit Label _____________________________________________
d
e
OFFSET
TERM
IMPED
DELAY
Z
s
Operating Concepts
Modifying Calibration Kits
SAVE USER KIT
Table
7-1.
FREQ (GHz)
COAX
or
WG
LOSS
MIN
MAX
0
/s
softkey, if
STND
LABEL
7- 59

Advertisement

Table of Contents
loading

This manual is also suitable for:

8753es

Table of Contents