Remove The Adapter - Agilent Technologies 8753ET User Manual

Network
Hide thumbs Also See for 8753ET:
Table of Contents

Advertisement

Remove the Adapter

When the two sets of error correction files have been created (now referred to as "calibration sets"), the A3
adapter may be removed.
1. Press
Cal
MORE
HELP ADAPT REMOVAL
RECALL CAL SETS
ADAPTER DELAY
ADAPTER COAX
ADAPTER WAVEGUIDE
REMOVE ADAPTER
2. Press
RECALL CAL SETS
RECALL CAL PORT 2
RECALL CAL SETS
NOTE
In the following two steps, calibration data is recalled, not instrument states.
3. From the disk directory, choose the file associated with the port 1 error correction, then press
RECALL CAL PORT 1
4. When this is complete, choose the file for the port 2 error correction and press
RECALL CAL PORT 2
5. When complete, press
6. Enter the value of adapter A3 electrical delay by pressing
value.
7. Select the appropriate key:
8. Press
REMOVE ADAPTER
and overwrite the current active calibration set in use.
This process uses up an internal memory register. The calibration in this register is not the calibration
created by adapter removal, rather it is a "scratch" calibration. You may wish to delete the register, or
re-save the new calibration in this register as shown in the following step.
9. To save the results of the new calibration set, press
INTERNAL MEMORY
NOTE
Adapter removal can leave a residual state in internal memory. This is not a valid instrument
state and should be deleted.
10. Connect the DUT to the network analyzer as shown in
ADAPTER REMOVAL
(This Help softkey provides a quick reference guide to using
the adapter removal technique.)
to display
RECALL CAL PORT 1
.
also displays the internal (or external if internal not used) disk file directory.
.
.
.
RETURN
ADAPTER COAX
to complete the technique for calculating the new error coefficients
RETURN
SAVE STATE
Calibrating for Increased Measurement Accuracy
Adapter Removal Using ECal (ES Analyzers Only)
to display the following menu:
and
ADAPTER DELAY
or
ADAPTER WAVEGUIDE
Save/Recall
SELECT DISK
.
Figure 6-30
to perform calibrated measurements.
and entering the
.
6- 73