Making Measurements
Using Test Sequencing
Loop counter decision making The analyzer has a numeric register called a loop counter. The value of this
register can be set by a sequence, and it can be incriminated or decremented each time a sequence repeats
itself. The decision making commands
and
IF LOOP COUNTER = 0
IF LOOP COUNTER <> 0
jump to another sequence if the stated condition is true. When entered into the sequence, these commands
require you to enter the destination sequence. Either command can jump to another sequence, or restart the
current sequence.
As explained in
"Embedding the Value of the Loop Counter in a Title" on page
1-108, the loop counter value
can be appended to a title. This allows customized titles for data printouts or for data files saved to disk.