8. To assign the calibration standards to the various TRL calibration classes, press:
Cal
CAL KIT
MODIFY
9. Since you previously designated standard #1 for the REFLECT standard, press:
1
x1
10. Since you previously designated standard #6 for the LINE/MATCH standard, press:
TRL LINE OR MATCH
11. Since you previously designated standard #4 for the THRU/LINE standard, press:
4
TRL THRU
12. To complete the specification of class assignments, press:
SPECIFY CLASS DONE
Label the Classes
NOTE
To enter the following label titles, an external keyboard may be used for convenience.
13. Press
LABEL CLASS
14. Change the label of the "TRL REFLECT" class to "TRLSHORT."
15. Change the label of the "TRL LINE OR MATCH" class to "TRLLINE."
16. Change the label of the "TRL THRU" class to "TRLTHRU."
17. Press
LABEL CLASS DONE
Label the Calibration Kit
18. Press
LABEL KIT
DONE.
19. To save the newly defined kit into nonvolatile memory, press:
KIT DONE (MODIFIED)
NOTE
Refer to
"Saving Modified Calibration Kits to a Disk" on page 7-65
saving modified calibration kits, along with calibration data and instrument states, to a disk.
Calibrating for Non-Coaxial Devices (ES Analyzers Only)
SPECIFY CLASS
6
x1
x1
MORE
MORE
.
and create a label up to 8 characters long. For this example, enter "TRL KIT1"
SAVE USER KIT
Calibrating for Increased Measurement Accuracy
MORE
MORE
.
TRL REFLECT
for information about
6- 51