7. After you connect the second ECal module, press
8. Repeat steps 4 and 5 if you selected to calibrate using the manual thru option.
9. Review the displayed calibration results. Refer to
10. Save the calibration results by pressing

Display the Module Information

Display information about the module by pressing
.
MODULE INFO
A text window is displayed that contains the following information about the selected module.
1
PC interface unit port to which it is connected
2
Model number
3
Serial number
4
Connector type
5
Date of last certification
Calibrating for Increased Measurement Accuracy
Calibrating Using Electronic Calibration (ECal)
CONTINUE ECal
"Perform the Confidence Check" on page
.
Save/Recall
ECal MENU
Cal
6
Number of measurement points
7
Start (minimum) frequency
8
Stop (maximum) frequency
9
Suggested warmup time
10
Warmup status (This is the remaining time on an
internal analyzer timer. The timer simply counts down
the recommended ECal module warmup time. When
the timer reaches zero, a message is displayed
indicating that the ECal module is ready. The ECal
module temperature or status is not read. The timer is
reset when the analyzer is powered up or preset or
when a new module is attached.)
to continue the ECal.
6-64.