Compensating For Directional Coupler Response - Agilent Technologies 8753ET User Manual

Network
Hide thumbs Also See for 8753ET:
Table of Contents

Advertisement

3. Press
and then press either the
EDIT
which part of the segment you want to edit.
❏ If you are modifying the frequency, enter the new value, followed by a
❏ If you are modifying the correction factor, enter the new value, followed by the
4. Press
after you have finished modifying the segment.
DONE
5. If you want to edit any other segments, press
step 2.
Deleting Frequency Segments
1. Access the Segment Modify menu by pressing
CAL FACTOR SENSOR A
is that you want to delete).
2. Identify the segment that you want to delete by pressing
keys to locate and position the segment next to the pointer (>), shown on the display. Or press
and enter the segment number followed by
SEGMENT
3. Press
.
DELETE
The analyzer deletes the segment and moves the remainder of the segments up one number.
4. You could also delete all the segments in a list by pressing
5. Press
when you are finished modifying the segment list.
DONE

Compensating for Directional Coupler Response

If you use a directional coupler to sample power in your measurement configuration, you should enter the
coupled arm power loss value into the power loss table, using the following procedure. You can enter the
loss information in a single segment, and the analyzer will assume that the value applies to the entire
frequency range of the instrument. Or, you can input actual measured power loss values at several
frequencies using up to 12 segments, enhancing power accuracy.
1. Press
PWRMTR CAL
Cal
The analyzer shows the notation EMPTY, if you have not entered any segment information.
2. To create the first segment, press
data point, followed by the appropriate key:
3. Press
and enter the power loss that corresponds to the attenuation of the directional coupler
LOSS
(or power splitter) at the frequency that you have entered in the previous step. Complete the power loss
entry by pressing
x1
NOTE
Remember to subtract the through arm loss from the coupler arm loss before entering it into
the power loss table, to ensure the correct power at the output of the coupler.
FREQUENCY
SEGMENT
Cal
(or
CAL FACTOR SENSOR B
LOSS/SENSR LISTS
ADD
FREQUENCY
G/n
.
DONE
Calibrating for Increased Measurement Accuracy
Power Meter Measurement Calibration
or
CAL FACTOR
G/n
and follow the previous steps, starting with
PWRMTR CAL
LOSS/SENSR LISTS
, depending on where the segment
and using the
SEGMENT
.
x1
CLEAR LIST
POWER LOSS
and enter a frequency of a correction factor
.
M/
k/m
key, depending on
,
, or
key.
M/
k/m
key.
x1
and
.
YES
.
6- 35