Frequency Offset Operation - Agilent Technologies 8753ET User Manual

Network
Hide thumbs Also See for 8753ET:
Table of Contents

Advertisement

Operating Concepts
Knowing the Instrument Modes
Figure
7-47.
Figure 7-47
Typical Test Setup for Tuned Receiver Mode
Tuned Receiver Mode In-Depth Description
If you press
System INSTRUMENT MODE
operates independently of any signal source.
The following features and limitations apply to the tuned receiver mode:
• It is a fully synthesized receiver; it does not phase-lock to any source.
• It functions in all sweep types.
• It requires a synthesized CW source whose timebase is input to the analyzer's external frequency
reference.
For more information on using the tuned receiver mode, refer to

Frequency Offset Operation

Refer to
"Conversion Loss Using the Frequency Offset Mode" on page 2-13
offset operation.
7- 84
TUNED RECEIVER
"Tuned Receiver Mode" on page
for information on frequency
, the analyzer receiver
2-26.