Making Measurements
Using Display Functions
Viewing Both Primary Measurement Channels
In some cases, you may want to view more than one measured parameter at a time. Simultaneous gain and
phase measurements, for example, are useful in evaluating stability in negative feedback amplifiers. You can
easily make such measurements using the dual channel display.
1. To see channels 1 and 2 in the same grid, press:
Display
DUAL | QUAD SETUP
1X.
Figure 1-7 Example of Viewing Channel 1 and 2 Simultaneously
2. To view the measurements on separate graticules, press: Set
channel 1 on the upper half of the display and channel 2 on the lower half of the display. The analyzer
defaults to measuring S
1-12
, set
DUAL CHAN on OFF
on channel 1 and S
on channel 2.
11
21
to ON, and
SPLIT DISP
to 2X. The analyzer shows
SPLIT DISP
to