Figure 2-29
Connections for the Second Portion of Conversion Compression Measurement
9. To set the frequency offset mode LO frequency, press:
System
INSTRUMENT MODE
LO MENU
FREQUENCY:CW
10. To select the converter type, press:
RETURN
UP CONVERTER
11. To select a low-side LO measurement configuration, press:
RF>LO
FREQ OFFS ON
In this low-side LO, up converter measurement, the analyzer source frequency is offset lower than the
receiver frequency. The analyzer source frequency can be determined from the following equation:
receiver frequency (800 MHz)  LO frequency (600 MHz) = 200 MHz
Conversion Compression Using the Frequency Offset Mode
FREQ OFFS MENU
600
M/
Making Mixer Measurements
2- 41