Making Measurements
Characterizing a Duplexer (ES Analyzers Only)
this example, a wide frequency range that covers both the Tx-Ant and Ant-Rx parameters has been
chosen.
4. Uncouple the primary channels from each other and then press
COUPLED CH on OFF
5. Press
CONFIGURE MENU
System
6. Set up the desired mode.
For K36 mode, toggle
For K39 mode, toggle
SELECT PORTS [1-3]
7. Perform a full two-port calibration on channel 1 (refer to
Measurement Accuracy,"
NOTE
Make sure you connect the standards to the Tx port of the test adapter (or a cable attached
to it) for FORWARD calibrations, and to the Ant port for REVERSE calibrations.
8. Save the instrument state by pressing
9. Press
.
Chan 2
10. Set up channel 2 for the Ant-Rx stimulus parameters. In this example, a wide frequency range that
covers both the Tx-Ant and Ant-Rx parameters has been chosen.
11. Set up control of the test adapter so that channels 2 and 4 are Rx:
• For K36 mode, press
• For K39 mode, press
12. Perform a full two-port calibration on channel 2.
NOTE
Make sure you connect the standards to the Rx port of the test adapter (or a cable attached
to it) for FORWARD calibrations, and to the Ant port for REVERSE calibrations.
13. Save this state in the analyzer:
Press
Save/Recall SAVE STATE
14. Connect the duplexer to the test adapter.
15. Set up a 2-graticule, 4-parameter display with transmission measurements on the top graticule and
reflection measurements on the bottom graticule:
Press
Display DUAL|QUAD SETUP
Trans: REV S12 (A/R)
Chan 4
Refl: REV S22 (B/R)
Refl: FWD S11 (A/R)
The display will be similar to
1-50
to
OFF
.
USER SETTINGS
K36 MODE on OFF
K39 MODE on OFF
.
if necessary).
Save/Recall SAVE STATE
SELECT [RX-ANT]
Meas
Meas
SELECT PORTS [2-3]
.
4-PARAM DISPLAYS
Chan 1
Trans:FWD S21 (B/R)
, then set
DUAL CHAN on OFF
Figure
1-41.
Sweep Setup
.
to ON. Then, press
Meas
to ON. Then, press
Meas
Chapter 6 , "Calibrating for Increased
.
.
.
SETUP B
to
ON
.
and toggle
SELECT [TX-ANT]