1. To access the ripple test menu, press:
System
LIMIT MENU
2. To access the ripple test edit menu, press
3. Add the first frequency band (Frequency Band 1) to be tested by pressing
4. Set the lower frequency value of Frequency Band 1 by pressing:
MINIMUM FREQUENCY
5. Set the upper frequency value of Frequency Band 1 by pressing:
MAXIMUM FREQUENCY
6. Set the maximum allowable ripple amplitude value of Frequency Band 1 by pressing:
MAXIMUM RIPPLE
7. Repeat steps 3 through 6 for the two remaining frequency bands to be tested for maximum ripple.
The network analyzer allows you to enter up to 12 frequency bands to be tested for maximum ripple.
8. After you have entered all of the ripple test frequency band parameters, return to the ripple test menu by
pressing
.
DONE

Editing Ripple Test Limits

Once the frequency band limits for ripple testing have been created, the limits may be changed using the
same menu that was used to create them. Using the edit ripple test menu, you may:
• Change existing frequency band limits
• Add more frequency band limits
• Delete individual frequency band limits
• Clear all frequency band limits
Changing Existing Frequency Band Limits
Existing frequency band limits may be changed for testing the ripple. This procedure guides you through
changing the existing frequency band limits.
1. To access the ripple test edit menu from the ripple test menu, press:
EDIT RIPL LIMIT
2. Enter the frequency band whose limits you want to change by pressing:
a.
FREQUENCY BAND
b. The numeric key indicating the frequency band number that you are changing.
The frequency band number is located in the left column of the list of frequency bands.
RIPPLE LIMIT
EDIT RIPL LIMIT

M/
3.2
G/n
2.0
x1
Making Measurements