Figure 1-74 Connections for a Bandpass Filter Example Measurement
2. Press
and choose the measurement settings. For this example, the measurement settings are
Preset
as follows:
a.
Meas
Trans: FWD S21 (B/R)
b.
Center
321
c.
Span
200
d.
AUTO SCALE
Scale Ref
You may also want to select settings for the number of data points, power, averaging, and IF bandwidth.
Figure 1-75 Filter Pass Band Before Bandwidth Test
3. Substitute a thru for the device and perform a response calibration by pressing:
or on ET models:
M/
M/
Using Bandwidth Limits to Test a Bandpass Filter
TRANSMISSN
Making Measurements