Perform The Confidence Check - Agilent Technologies 8753ET User Manual

Network
Hide thumbs Also See for 8753ET:
Table of Contents

Advertisement

Calibrating for Increased Measurement Accuracy
Calibrating Using Electronic Calibration (ECal)

Perform the Confidence Check

The confidence check is a means of visually checking the quality of the calibration. The confidence check
displays the currently measured data (DATA trace) and the factory-premeasured data (MEM trace) for the
module's confidence state. The confidence state, an independent reference standard, was not used for the
calibration. For a good calibration measurement, there should be no significant difference between the
traces (see
Figure
6-24).
To Perform the Confidence Check
NOTE
A confidence check is only valid after an ECal has been performed.
1. Press
Cal
ECal MENU
2. Press
MODULE A b
3. Select
CONFIDENCE CHECK
4. Press
PARAMETER [ ]
Pressing the
PARAMETER [ ]
models, TRANS (transmission) and REFL (reflection)) calibration traces. The confidence check will only
display the S-parameter calibration data for which you calibrated. Uncalibrated S-parameter data traces
are invalid. The memory trace still displays the module's factory-premeasured trace.
For example,
If you performed:
A 1-port calibration
S11 1-PORT
(ES models)
S22 1-PORT
(ES models)
REFLECTION 1-PORT
An enhanced response calibration
S11/S21 ENH. RESP.
S22/S12 ENH. RESP.
TRANS/REFL ENH. RESP.
A full 2-port calibration
FULL 2-PORT
5. Press
TRACE TYPE [ ]
displayed.
Pressing the
TRACE TYPE [ ]
confidence check can display the measured ECal results (DATA) and the premeasured calibration data
(MEM) in following five ways.
6-64
CONFIGURE [ ]
until the correct module (A or B) is selected.
to display the confidence check and the ECal Confidence Menu.
until the S-parameter that you want to view is displayed.
softkey toggles between the S11, S21, S12, and S22 (or for the ET
(ET models)
(ES models)
(ES models)
(ET models)
(ES models)
until the calibration confidence check trace that you want to view is
softkey toggles between the five trace-type display options. The
.
You can view:
Only one calibrated S-parameter
Only two calibrated S-parameters, one at a
time
All four calibrated S-parameters, one at a time