Agilent Technologies 8753ET User Manual page 90

Network
Hide thumbs Also See for 8753ET:
Table of Contents

Advertisement

Making Measurements
Using Limit Lines to Test a Device
5. To terminate the flat line segment by establishing a single point limit, press:
ADD
STIMULUS VALUE
LIMIT TYPE
SINGLE POINT
Figure 1-61
shows the flat limit lines that you have just created with the following parameters:
• stimulus from 127 MHz to 140 MHz
• upper limit of 21 dB
• lower limit of 27 dB
Figure 1-61 Example Flat Limit Line
• To create a limit line that tests the low side of the filter, press:
ADD
STIMULUS VALUE
UPPER LIMIT
LOWER LIMIT
DONE
LIMIT TYPE
ADD
STIMULUS VALUE
DONE
LIMIT TYPE
1-76
140
M/
RETURN
100
M/
65
x1
200
x1
FLAT LINE
RETURN
123
M/
SINGLE POINT
RETURN
DONE

Advertisement

Table of Contents
loading

This manual is also suitable for:

8753es

Table of Contents