Making Mixer Measurements
Phase or Group Delay Measurements
Figure 2-24
Connections for a Group Delay Measurement
6. To select the converter type and a high-side LO measurement configuration, press:
System
INSTRUMENT MODE
DOWN CONVERTER
7. To view the measurement results on the analyzer's display, press:
System
INSTRUMENT MODE
8. To select the format type, press:
Format
DELAY
9. To make a response error-correction, press:
INPUT PORTS
Meas
Cal
CALIBRATE MENU
2-36
FREQ OFFS MENU
RF<LO
FREQ OFFS ON
FREQ OFFS MENU
B/R
RESPONSE
THRU
VIEW MEASURE