4. Reconnect your test device.
5. To better view the measurement trace, press:
Scale Ref
AUTO SCALE

Creating Flat Limit Lines

In this example procedure, the following flat limit line values are set:
Frequency Range
127 MHz to 140 MHz
100 MHz to 123 MHz
146 MHz to 160 MHz
The minimum value for measured data is 200 dB.
NOTE
1. To access the limits menu and activate the limit lines, press:
System
LIMIT MENU
CLEAR LIST
YES
2. To create a new limit line, press:
ADD
The analyzer generates a new segment that appears on the center of the display.
3. To specify the limit's stimulus value, test limits (upper and lower), and the limit type, press:
STIMULUS VALUE
UPPER LIMIT
LOWER LIMIT
NOTE
You could also set the upper and lower limits by using the
DELTA LIMITS
MIDDLE VALUE
This would correspond to a test specification of 24 3 dB.
4. To define the limit as a flat line, press:
LIMIT TYPE
FLAT LINE
Power Range
27 dB to 21 dB
200 dB to 65 dB
200 dB to 65 dB
LIMIT LINE
127
M/
21
x1
27
x1
DONE
keys. To use these keys for the entry, press:
24
x1
RETURN
Using Limit Lines to Test a Device
LIMIT LINE ON
EDIT LIMIT LINE
MIDDLE VALUE
DELTA LIMITS
3
Making Measurements
and
x1
1- 75