Making Measurements
Using Limit Lines to Test a Device
1. To access the limits menu and activate the limit lines, press:
System
LIMIT MENU
CLEAR LIST
YES
2. To establish the start frequency and limits for a sloping limit line that tests the low side of the filter,
press:
ADD
STIMULUS VALUE
UPPER LIMIT
DONE
LIMIT TYPE
3. To terminate the lines and create a sloping limit line, press:
ADD
STIMULUS VALUE
UPPER LIMIT
DONE
LIMIT TYPE
4. To establish the start frequency and limits for a sloping limit line that tests the high side of the filter,
press:
ADD
STIMULUS VALUE
UPPER LIMIT
DONE
LIMIT TYPE
5. To terminate the lines and create a sloping limit line, press:
ADD
STIMULUS VALUE
UPPER LIMIT
DONE
LIMIT TYPE
You could use this type of limit to test the shape factor of a filter.
1-78
LIMIT LINE
123
M/
65
x1
LOWER LIMIT
SLOPING LINE
125
M/
26
x1
LOWER LIMIT
SINGLE POINT
144
M/
26
x1
LOWER LIMIT
SLOPING LINE
146
M/
65
LOWER LIMIT
x1
SINGLE POINT
LIMIT LINE ON
EDIT LIMIT LINE
200
x1
RETURN
200
x1
RETURN
200
x1
RETURN
200
x1
RETURN