1. Set the LO source to the desired CW frequency of 1000 MHz and power level to 13 dBm.
2. Initialize the analyzer by pressing
3. Set the analyzer's LO frequency to match the frequency of the LO source by pressing:
System
INSTRUMENT MODE
FREQUENCY: CW
4. From the front panel of the analyzer, set the desired receiver frequency and source output power by
pressing:
Center
300
Span
100
PWR RANGE MAN
Power
0
5. Connect the instruments as shown in
between PORT 1 and PORT 2.
CAUTION
To prevent connector damage, use an adapter (part number 1250-1462) as a connector saver
for R CHANNEL IN.
Useful Frequency Range
0.03 to 3 GHz
dc to 1250 MHz
.
Preset
FREQ OFFS MENU
1000
M/
M/
M/
x1
Figure
2-24, placing a broadband "calibration" mixer (ZFM-4)
Making Mixer Measurements
Phase or Group Delay Measurements
Group Delay
0.5 ns
0.6 ns
LO MENU
2- 35