Removing a TRIZAIC source

You can remove the TRIZAIC UPLC source and replace it with a conventional
nanoflow interface on SYNAPT-series or Xevo-series mass spectrometers.
"Installing the TRIZAIC UPLC source" on page
See also:
Required materials
Chemical-resistant, powder-free gloves
¼-inch wrench
Screwdriver
To remove the TRIZAIC source:
Warning:
with biohazardous and/or toxic materials. Always wear
chemical-resistant, powder-free gloves while performing this
procedure.
Warning:
performed on its source before commencing this procedure.
1.
Prepare the instrument for working on its source (see
Warning:
take great care while working with these components.
2.
Remove the nanoTile from the docking port (see
3.
Shut down the MassLynx software.
4.
Power-off the nanoACQUITY UPLC sample manager.
5.
Disconnect the QSPI cable.
6.
Using the ¼-inch wrench, loosen and disconnect the fluid waste line and
fluid inlet lines from the sample manager (see
connections:" on page
7.
Swing open the TRIZAIC source enclosure unit from the source
mounting on the mass spectrometer.
The probe and source components can be contaminated
To avoid electric shock, prepare the instrument for work
The probe and source can be hot. To avoid burn injuries,
2-9).
2-1.
page
page
2-19).
figure "Plumbing
Removing a TRIZAIC source
2-1).
3-9