2.
In the Instrument Console, click Standby
illuminated.
3.
Wait 3 minutes to allow the desolvation gas flow to cool the probe and source.
4.
In the Instrument Console, click API
5.
Lift the visor on the front of the instrument so that it is clear of all the source components
and probe.
4.6

Removing and refitting the source enclosure

4.6.1

Removing the source enclosure from the instrument

4.6.1.1
Required material
Chemical-resistant, powder-free gloves
Warning:
To avoid personal contamination with biologically hazardous, or toxic materials, and to
avoid spreading contamination to uncontaminated surfaces, wear clean, chemical-resistant,
powder-free gloves when working with the source components.
To remove the source enclosure:
1.
Prepare the instrument according to the procedure on
for operations on or inside its
Warning:
source; they can be hot.
2.
Remove the ESI probe from the source (see
3.
Disconnect the probe adjuster and options cables from the instrument's connectors.
4.
Pull the source enclosure release (located at the bottom, right-hand side) outward, and
swing open the enclosure.
source".
To avoid burn injuries, take great care while working with the probe and
September 7, 2016, 715003560IVD Rev. C
Page 45
, and confirm that the Operate indicator is not
, to stop the desolvation gas flow.
page
44, "To prepare the instrument
page
39).