Removing and refitting the source enclosure

Removing the source enclosure from the instrument

Required materials
Chemical-resistant, powder-free gloves
To remove the source enclosure:
Warning:
biohazardous and/or toxic materials. Always wear
chemical-resistant, powder-free gloves while performing this
procedure.
1.
Prepare the instrument for working on the source (see
Warning:
take great care while working with these components.
2.
Remove the ESI probe from the source (see
3.
Slide open the instrument's source interface door (see the figure on
page
1-4).
4.
Disconnect the probe adjuster and options cables from the instrument's
connectors.
5.
Pull the source enclosure release (located at the bottom, right-hand side)
outwards, and swing open the enclosure.
The source components can be contaminated with
The probe and source can be hot. To avoid burn injuries,
Removing and refitting the source enclosure
page 4-6).
page
3-6).
4-7