Removal and refitting of the source enclosure

Removing the source enclosure from the instrument

Required material
Chemical-resistant, powder-free gloves
To remove the source enclosure
1.
Prepare the instrument for working on the source (see
instrument for working on the source" on page 4-7).
2.
Remove the probe from the source, see
page
3-6.
3.
Slide open the instrument's source interface door (see the figure on
page
1-4).
4.
Disconnect the probe adjuster and options cables from the instrument's
connectors.
5.
Pull the source enclosure release (located at the bottom, right-hand side)
outwards and swing open the enclosure.
4-8
Maintenance Procedures
The source components can be contaminated with
Warning:
biohazardous and/or toxic materials. Always wear
chemical-resistant, powder-free gloves while performing this
procedure.
The probe and source can be hot. To avoid burn
Warning:
injuries, take great care while working with these components.
"Preparing the
"Removing the ESI probe" on