3 Changing the Mode of Operation
7.
Connect the HT cable to the instrument's front panel connector.
Notice:
ensure that the lamp assembly does not touch the corona pin when
the source enclosure door is closed.
8.
Install the IonSABRE II probe to the source, and ensure that it is
working correctly (see
An automatic pressure test runs each time you close the source
Tip:
enclosure and when the instrument starts.
Removing the IonSABRE II probe and APPI/APCI source
enclosure
Required material
Chemical-resistant, powder-free gloves
Warning:
toxic materials, and to avoid spreading contamination to
uncontaminated surfaces, wear clean, chemical-resistant,
powder-free gloves while performing this procedure. The source
components can be contaminated.
Warning:
prepare the instrument according to the procedure on
"Preparing the instrument for operations on or inside its
To remove the combined APPI/APCI source
1.
Prepare the instrument for working on the source (see
Warning:
with the probe and source; these components can be hot.
2.
Remove the IonSABRE II probe (see
3.
Disconnect the HT cable from the instrument's front panel.
4.
Disconnect the APPI drive cable from the instrument's front panel.
5.
Remove the source enclosure (see
72
March 14, 2014, 715004387 Rev. B
To prevent damage to the corona pin and lamp assembly,
page
63).
To avoid personal contamination with biohazards or
To avoid electric shock, before beginning this procedure,
To avoid burn injuries, take great care while working
:
page
65).
page
90).
page
88,
source".
page 88).