Removing The Ionsabre Ii Probe And Combined Appi/Apci Source Enclosure - Waters Xevo TQ-S Operator's, Overview And Maintenance Manual

Mass spectrometry system
Table of Contents

Advertisement

9.
Slide closed the instrument's source interface door.
10. Install the IonSABRE II probe to the source, and ensure that it is
working correctly (see
An automatic pressure test is performed each time the source
Tip:
enclosure is closed, and when the instrument starts.
Removing the IonSABRE II probe and combined APPI/APCI
source enclosure
Required materials
Chemical-resistant, powder-free gloves
To remove the combined APPI/APCI source
Warning:
prepared before commencing this procedure.
1.
Prepare the instrument for working on the source (see
2.
Remove the IonSABRE II probe (see
3.
Disconnect the HT cable, from the instrument's front panel.
4.
Disconnect the APPI drive cable, from the instrument's front panel.
5.
Remove the source enclosure (see
6.
Remove the corona pin (see
7.
Fit the blanking plug to the pin's mounting contact.
page
The source components can be contaminated with
Warning:
biohazardous and/or toxic materials. Always wear
chemical-resistant, powder-free gloves while performing this
procedure.
To avoid electric shock, ensure that the instrument is suitably
The probe and source can be hot. To avoid burn injuries,
Warning:
take great care while working with these components.
3-8).
:
page
3-10).
page
4-9).
page
4-15).
Combined APPI/APCI source
page 4-8).
3-17