Removing the Combined APPI/APCI Source

Removing the APCI probe and APPI lamp drive assembly

To complete this procedure, you must perform a pressure test on the
Note:
source, as described in the Waters Source Pressure Test Unit Operator's Guide.
Required materials
Chemical-resistant, powder-free gloves
Needle-nose pliers
To remove the APCI probe and APPI lamp drive assembly from the
instrument
1.
On the MassLynx Tune window, click Press for Standby, and confirm
that the instrument status indicator is red.
2.
Wait for three minutes for the desolvation gas to cool the probe.
3.
Click
4.
Remove the APCI probe from the source.
See also:
Guide.
2-8
The Combined APPI/APCI Source
to stop the nitrogen flow.
To avoid self -contamination with toxic and
Warning:
biohazardous materials, always wear chemical-resistant,
powder-free gloves at all times while handling the
components.
To avoid electric shock, ensure that the instrument is
Warning:
in standby while working with the instrument's front access door
open.
To avoid burn injuries, take great care while working
Warning:
with the instrument's front access door open.The probe and
source may be hot.